Adaptive Filtering and Spectral Analysis of Pulsed Doppler Ultrasound to Aid the Assessment of Cardiovascular Condition

机译:脉冲多普勒超声的自适应滤波和频谱分析有助于评估心血管疾病

获取原文

摘要

This paper presents a new approach to filtering wall motion artefact from Doppler ultrasound signals. The technique tracks the wall motion and removes it from the backscattered signal, thus preserving low frequency blood flow information, which would have otherwise been removed had conventional highpass filtering methods been applied. This provides a much-improved representation of the blood flow profile throughout the cardiac cycle. Often when cardiovascular vessels are scanned the Doppler SNR is typically OdB due to the attenuation of tissue, fat and general in-homogenates in the separating medium. The effects of this speckle are much more emphasised after wall motion is removed, as the noise floor is lifted. Hence this paper also examines four adaptive filtering techniques for minimising the effects of this noise in the backscattered Doppler ultrasound signal.
机译:本文提出了一种从多普勒超声信号中滤除壁运动伪影的新方法。该技术跟踪壁运动并将其从后向散射信号中去除,从而保留了低频血流信息,否则,如果应用常规的高通滤波方法,该信息将被去除。这提供了整个心动周期中血流曲线的大大改善的表示。通常,由于分离介质中组织,脂肪和一般均质物质的衰减,扫描心血管血管时,多普勒SNR通常为OdB。在消除壁运动之后,由于本底噪声被提升,因此斑点的影响更加突出。因此,本文还研究了四种自适应滤波技术,以将这种噪声对后向散射多普勒超声信号的影响降至最低。