摘要

In the presence of metal implant, it can induce streak artefacts appearance in computed tomography (CT) images that can degrade image quality and lead to misdiagnosis. A comparative study of two different sinogram interpolation methods for metal artefact reduction in CT images is presented. The metal part of the sinogram was determined and removed by thresholding technique. The missing part of the metal trace on the sinogram was interpolated by two interpolation methods which were cubic spline and Laplace interpolation. Both methods were applied to the selected CT images of different metal geometry and complexity. The effectiveness of both interpolation methods was compared. The results showed Laplace interpolation method produced the better results of gap filling as compared to cubic spline interpolation. But, it produces lower signal-to-noise ratio (SNR) values.
机译:在金属植入物的存在下,它会导致计算机断层扫描(CT)图像中出现条纹伪影,从而降低图像质量并导致误诊。提出了两种减少CT图像中金属伪影的正弦图插值方法的比较研究。通过阈值技术确定并去除了正弦图的金属部分。通过两种插值方法(三次样条和拉普拉斯插值)对正弦图上金属迹线的缺失部分进行插值。两种方法都应用于不同金属几何形状和复杂度的所选CT图像。比较了两种插值方法的有效性。结果表明,与三次样条插值相比,拉普拉斯插值方法产生的间隙填充效果更好。但是,它会产生较低的信噪比(SNR)值。