Effect of tubing length pump on flow derived index during rotary blood pump support

机译:旋转式血泵支撑期间管长泵对流量衍生指数的影响

获取原文

摘要

The variable of connecting tube length (CTL) in rotary blood pumps (RBPs) is normally existed for the external circulatory support and thus eventually incurs the different outcome as compared to the internal circulatory support. This study aims to determine the effect of the various CTL on the measured parameters from RBP. To serve this purpose, a derived indexes from RBP under mock circulatory loop (MCL) were examined. IQ which is a slope of the linear regression between the maximal time derivative of pump flow, dQ/dtmax, and its peak to peak, QP2P, was determined to evaluate the effect of 3 different CTLs; 130, 50 and 10 cm (length of inlet tube), and 130, 50 and 50 cm (length of outlet tube). Moreover, the parameters dH/dtmax, which are the maximal time derivative of pressure head (H); AOP-LVP, and its peak to peak were derived. The slope of a linear regression between dQ/dtmax and dH/dtmax decreased during reduce CTL in contrast to IQ index. Thus, this study proved that IQ index was not significant influenced by a hydraulic resistance (including CTL) and therefore can be used to monitor the cardiac function in case of both external and internal RBP support.
机译:旋转血液泵(RBP)中的连接管长度(CTL)变量通常存在于外部循环支架中,因此与内部循环支架相比最终会产生不同的结果。本研究旨在确定各种CTL对RBP测得参数的影响。为了达到这个目的,在模拟循环回路(MCL)下检查了从RBP导出的索引。确定IQ是泵流量的最大时间导数dQ / dtmax及其峰到峰QP2P之间的线性回归斜率,以评估3种不同CTL的效果; 130、50和10厘米(入口管的长度)以及130、50和50厘米(出口管的长度)。此外,参数dH / dtmax是压力头(H)的最大时间导数;推导了AOP-LVP及其峰到峰。与IQ指数相比,在降低CTL期间dQ / dtmax和dH / dtmax之间的线性回归斜率减小。因此,这项研究证明,IQ指数不受液压阻力(包括CTL)的影响不明显,因此可以在外部和内部RBP支持的情况下用于监测心功能。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号