Background: Many factors can influence the radiant power delivered by the low-level laser therapy (LLLT) equipment, such as its cleaning and condition, as well as the use of plastic films for protecting the laser or even its time of use. Radiant power is an important factor to consider because it affects the amount of energy delivered to the target tissue. The difference between real radiant power (RRP) and nominal radiant power (NRP) may interfere in the expected results, because the delivered energy is different from the desired energy. Purpose: The objective of this study was to compare the NRP with the RRP offered by LLLT devices under clinical conditions of use. Material and Methods: For data collection to this study, 61 LLLT devices used in private and public dental practices in the state of Tocantins, Brazil, were evaluated. Three consecutive power measurements were performed at one-minute intervals and then the average of the measured power was calculated. The RRP was compared to the NRP. Results: The equipment presented NRP from 30 to 500mW while RRP ranged from 17.3 to 107.0mW. Discussion and Conclusion: The mean power measured in clinical conditions of use of the laser equipment was different from the nominal power reported by the manufacturers of the devices (p<0,01). The RRP ranged between 12.92% and 107% of NRP. This fact is worrisome, since one of the most important parameters for the success of the treatment of an injury using LLLT is the energy (power x time) delivered. These findings reinforce the need of calibrating the equipment before each laser application in order to avoid failures in the therapeutic conduct.
机译:背景:许多因素可能会影响低水平激光治疗(LLLT)设备传递的辐射功率,例如其清洁和状况,以及使用塑料膜来保护激光甚至使用时间。辐射功率是要考虑的重要因素,因为它会影响传递到目标组织的能量。实际辐射功率(RRP)与标称辐射功率(NRP)之间的差异可能会干扰预期结果,因为所传递的能量不同于所需的能量。目的:本研究的目的是比较NLT与LLLT设备在临床使用条件下提供的RRP。材料和方法:为了收集该研究的数据,对巴西Tocantins州用于私人和公共牙科诊所的61种LLLT设备进行了评估。以一分钟的间隔执行三个连续的功率测量,然后计算测量功率的平均值。将RRP与NRP进行比较。结果:设备的NRP从30到500mW,而RRP从17.3到107.0mW。讨论与结论:在使用激光设备的临床条件下测得的平均功率不同于设备制造商报告的标称功率(p <0.01)。 RRP介于NRP的12.92%至107%之间。这个事实令人担忧,因为使用LLLT成功治疗伤害最重要的参数之一就是传递的能量(功率x时间)。这些发现增加了在每次使用激光之前校准设备的必要性,以避免治疗行为的失败。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号