首页> 外文期刊>biotechnology progress >Practical Considerations in the Measurement of Culture Fluorescence
【24h】

Practical Considerations in the Measurement of Culture Fluorescence

机译:Practical Considerations in the Measurement of Culture Fluorescence

获取原文
           

摘要

AbstractWe have examined practical considerations associated with the use of a commercially available fluorescence probe for in situ measurements in bioreactors. The optical path length of the measurement was first determined and a flow cell subsequently designed. The environment (agitation/aeration rates) of the probe was found to have a significant influence on the measurement. These effects were eliminated by placing the probe in a recycle loop using the flow cell. Fluorescence measurement in the recycle loop was verified to be representative of the cell sample and to not affect cell metabolism.

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com