Patient-specific desktop 3D-printed guides for pelvic tumour resection surgery: a precision study on cadavers

机译:Patient-specific desktop 3D-printed guides for pelvic tumour resection surgery: a precision study on cadavers

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose 3D-printed patient-specific instruments have become a useful tool to improve accuracy in pelvic tumour resections. However, their correct placement can be challenging in some regions due to the morphology of the bone, so it is essential to be aware of the possible placement errors in each region. In this study, we characterize these errors in common pelvic osteotomies. Methods We conducted an experiment with 9 cadaveric specimens, for which we acquired a pre-operative computed tomography scan. Small PSIs were designed for each case following a realistic surgical approach for four regions of the pelvis: iliac crest (C), supra-acetabular (S), ischial (I), and pubic (P). Final surgical placement was based on a post-operative scan. The resulting positions were compared with pre-operative planning, obtaining translations, rotations, and maximum osteotomy deviations in a local reference frame defined based on the bone's morphology. Results Mean translations and rotations in the direction of the osteotomy plane were as follows: C= 5.3 mm, 6.7°; S = 1.8 mm, 5.1°; I= 1.5 mm, 3.4°; P = 1.8 mm, 3.5°. Mean translations in the remaining axes were below 2 mm. Maximum osteotomy deviations (75% of cases) were below 11.8 mm in C(7.8 mm for half-length), 7.8 mm in S (5.5 mm for half-length), 5.5 mm in I, and 3.7 mm in P. Conclusion We have characterized placement errors for small PSIs in four regions of the pelvis. Our results show high errors in C and S PSIs in the direction of the resection plane's normal, and thus large osteotomy deviations. Deviations in short osteotomies in S, I and P and placement errors in the remaining directions were low. The PSIs used in this study are biocompatible and can be produced with a desktop 3D printer, thus minimizing manufacturing cost.

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号