首页> 外文期刊>Applied optics >Time-resolved imaging of solid phantoms for optical mammography
【24h】

Time-resolved imaging of solid phantoms for optical mammography

机译:固体模体的时间分辨成像,用于乳腺X线摄影

获取原文
获取原文并翻译 | 示例
           

摘要

We have recorded time-resolved transillumination images of solid phantoms with objects embedded that differ in their scattering and absorption coefficients from those of the bulk material, simulating a compressed human breast with a tumor inside. Employing time-correlated single photon counting at rates of up to 1 MHz, we recorded distributions of times of flight of photons at 1369 scan positions within 2.5 min. Several quantities, such as fractional transmittance, first moments, Fourier amplitudes, phase shifts, and frequency-dependent effective transport scattering and absorption coefficients, have been derived from experimental data to form two-dimensional images. By recording such images at a selected total number of photons detected, we have determined the contrast and effective signal-to-noise ratio in each case. (C) 1997 Optical Society of America
机译:我们已经记录了固体幻像的时间分辨透射照明图像,其中嵌入的物体的散射和吸收系数与散装材料的散射和吸收系数不同,模拟了内部有肿瘤的受压人类乳房。使用与时间相关的单个光子,计数速率高达1 MHz,我们记录了2.5分钟内1369个扫描位置的光子飞行时间分布。已经从实验数据中得出了几个数值,例如分数透射率,第一矩,傅立叶振幅,相移以及依赖于频率的有效传输散射和吸收系数,以形成二维图像。通过以选定的检测到的光子总数记录此类图像,我们确定了每种情况下的对比度和有效信噪比。 (C)1997年美国眼镜学会