We evaluated the use of a commercial flatbed scanner for digitizing photographic plates used for spectroscopy. The scanner has a bed size of 420 mm by 310 mm and a pixel size of about 0.0106 mm. Our tests show that the closest line pairs that can be resolved with the scanner are 0.024 mm apart, only slightly larger than the Nyquist resolution of 0.021 mm expected from the 0.0106 mm pixel size. We measured periodic errors in the scanner using both a calibrated length scale and a photographic plate. We find no noticeable periodic errors in the direction parallel to the linear detector in the scanner, but we find errors with an amplitude of 0.03 to 0.05 mm in the direction perpendicular to the detector. We conclude that large periodic errors in measurements of spectroscopic plates using flatbed scanners can be eliminated by placing the plate along the short side of the scanner and scanning the plates with the dispersion direction parallel to the linear detector.
机译:我们评估了商业平板扫描仪的使用,用于数字化用于光谱学的摄影板。 扫描仪的床尺寸为420毫米310毫米,像素尺寸为约0.0106毫米。 我们的测试表明,可以用扫描仪解决的最接近的线对分开为0.024 mm,仅略大于0.0106毫米像素尺寸的0.021mm的奈奎斯特分辨率。 我们使用校准的长度和照相板测量扫描仪中的定期误差。 我们在平行于扫描仪中的线性检测器的方向上发现没有明显的周期性误差,但在垂直于检测器的方向上发现幅度为0.03至0.05mm的误差。 我们得出结论,通过将板沿扫描仪的短侧放置并扫描平行于线性检测器的分散方向,可以消除使用平板扫描仪的光谱板测量中的大规模周期性误差。