Multibeam long-path differential optical absorption spectroscopy instrument: a device for simultaneous measurements along multiple light paths

机译:多光束长径差分光学吸收光谱仪:一种用于同时测量多个光路的设备

获取原文
获取原文并翻译 | 示例
           

摘要

A novel long-path differential optical absorption spectroscopy (DOAS) apparatus for measuring tropo-spheric trace gases and the first results from its use are presented: We call it the multibeam instrument. It is the first active DOAS device that emits several light beams simultaneously through only one telescope and with only one lamp as a light source, allowing simultaneous measurement along multiple light paths. In contrast to conventional DOAS instruments, several small mirrors are positioned near the lamp, creating multiple virtual light sources that emit one light beam each in one specific direction. The possibility of error due to scattering between the light beams is negligible. The trace-gas detection limits of NO_(2), SO_(2), O_(3), and H_(2)CO are similar to those of the traditional long-path DOAS instrument.
机译:提出了一种新颖的用于测量对流层痕量气体的长距离差分光学吸收光谱仪(DOAS),并提出了使用它的最初结果:我们称其为多光束仪器。它是第一台主动式DOAS设备,仅通过一个望远镜并仅以一个灯作为光源即可同时发射多个光束,从而可以沿多个光路同时进行测量。与传统的DOAS仪器相比,在灯附近放置了几个小镜子,创建了多个虚拟光源,每个虚拟光源在一个特定方向上发射一个光束。由光束之间的散射引起的误差的可能性可以忽略不计。 NO_(2),SO_(2),O_(3)和H_(2)CO的痕量气体检测极限与传统的长距离DOAS仪器相似。