The potential for retinal photic injury from exposure to endoilluminators was evaluated. The spectral irradiance for each endoilluminator configuration was weighted with the American Conference of Government Industrial Hygienists (ACGIH) aphakic action spectrum. The result was compared with the threshold limit value (TLV) published by the ACGIH and a time to TLV (time_(TLV)) was calculated for each configuration. The calculated time_(TLV) ranged from 0.27 to 3.5 min, times that are significantly shorter than typical operating times. The effects of incorporating short-wavelength cutoff filters were evaluated and found to significantly increase the time_(TLV). Exposure reduction techniques for use during surgery are discussed.
机译:评估了暴露于内照射器对视网膜造成的光损伤的可能性。使用美国政府工业卫生学家会议(ACGIH)无晶状体运动光谱对每种内照明器配置的光谱辐照度进行加权。将结果与ACGIH发布的阈值极限值(TLV)进行比较,并针对每种配置计算出到达TLV的时间(time_(TLV))。计算的time_(TLV)在0.27到3.5分钟之间,这比典型的操作时间要短得多。评估了合并短波长截止滤波器的效果,发现该效果大大增加了time_(TLV)。讨论了手术中减少接触的技术。