首页> 外文期刊>Iranian journal of science and technology >Calculating the Medium Correction Factor for the PTW-30013 Ionization Chamber
【24h】

Calculating the Medium Correction Factor for the PTW-30013 Ionization Chamber

机译:计算PTW-30013电离室的中等校正因子

获取原文
获取原文并翻译 | 示例
           

摘要

In this paper, the medium correction factor for lung, muscle, blood and bone was calculated using analytical method and MCNP simulation for different beam qualities. To validate the calculation method, the medium correction factor of the PTW-30013 ionization chamber for air was specified by measuring the air kerma and the absorbed dose to water calibration coefficients in the standard radiation field of 60Co in Secondary Standards Dosimetry Laboratory (SSDL) of Iran. The value of the medium correction factor for muscle and blood was gotten close to one because of their close density and atomic composition to water. For lung, the medium correction factor was calculated more over 1.08 for different beam qualities. Also, this factor for the bone was determined less than 0.98. The results of this study revealed that for accurate measurement of absorbed dose in low (e.g., lung) or high (e.g., bone) density materials, by ionization chambers calibrated in term of absorbed dose to water, the medium correction factor has to be considered.
机译:本文使用分析方法和MCNP模拟计算了不同光束素质的肺,肌肉,血液和骨骼的中等校正因子。为了验证计算方法,通过在二级标准剂量测定实验室(SSDL)中的60Co的标准辐射场中的空气校准和吸收剂量来指定用于空气的PTW-30013电离室的中等校正因子。伊朗。由于它们的密度和原子组合物与水的密度和原子组合物,肌肉和血液的培养基校正因子的值被接近。对于肺,介质校正因子以不同的光束质量计算更多1.08。而且,该骨的该因子测定小于0.98。该研究的结果表明,对于低(例如,肺)或高(例如,骨)密度材料的吸收剂量的精确测量,通过在吸收剂量的术语中校准的电离室,必须考虑介质校正因子。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号