首页> 外文期刊>IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control >3-D Ultrasound Imaging Using Helicoid Array Transducers
【24h】

3-D Ultrasound Imaging Using Helicoid Array Transducers

机译:3-D使用螺旋阵列换能器的超声成像

获取原文
获取原文并翻译 | 示例
           

摘要

Real-time 3-D intracardiac echocardiography (ICE) can enable faster imaging of surfaces orthogonal to the transducer, such as the pulmonary vein (PV) antra and cardiac valve annuli. However, the requirement for a 2-D grid of individually wired elements makes a traditional matrix array challenging to implement within an intravenous catheter. Helicoid array transducers are linear array transducers twisted about their long axis, allowing imaging of different elevation slices using sub-apertures. In this work, we examined the 3-D imaging characteristics of helicoid array transducers through simulations using Field II software and experimental measurements. We report results for varying transducer parameters, such as twist rate and sub-aperture size. We also discuss design considerations for these imaging parameters as they pertain to volumetric imaging of the heart.
机译:实时3-D心内超声心动图(ICE)可以更快地成像与换能器正交的表面,例如肺静脉(PV)ANTRA和心脏瓣膜载载。然而,单独有线元件的2-D网格的要求使传统的矩阵阵列具有挑战性地在静脉内导管内实现。螺旋阵列换能器是围绕它们的长轴扭曲的线性阵列换能器,允许使用子孔的不同仰卧片的成像。在这项工作中,我们通过使用Field II软件和实验测量来检查螺旋阵列换能器的3-D成像特性。我们报告结果改变变换器参数,例如扭速和子光圈尺寸。我们还讨论了这些成像参数的设计考虑因素,因为它们属于心脏的体积成像。