Ultrasound can be delivered transcranially to ablate brain tissue, open the blood-brain barrier, or affect neural activity. Transcranial focused ultrasound in small rodents is typically done with low- frequency single-element transducers, which results in unspecific targeting and impedes the concurrent use of fast neuroimaging methods. In this article, we devised a wide-angle spherical array bidirectional interface for high-resolution parallelized optoacoustic imaging and transcranial ultrasound (POTUS) delivery in the same target regions. The system operates between 3 and 9 MHz, allowing to generate and steer focal spots with widths down to 130 mu m across a field of view covering the entire mouse brain, while the same array is used to capture high-resolution 3-D optoacoustic data in real time. We showcasethe system'sversatilebeam-forming capacities aswell as volumetric optoacoustic imaging capabilities and discuss its potential to noninvasively monitor brain activity and various effects of ultrasound emission.
机译:超声可以通过经晶体递送至消融脑组织,打开血脑屏障,或影响神经活动。小型啮齿动物中的经颅聚焦超声通常用低频单元素换能器进行,这导致非特异性靶向并阻碍了快速神经影像方法的同时使用。在本文中,我们设计了一个广角球形阵列双向接口,用于在同一目标区域中的高分辨率并行光声成像和经颅超声(Potus)输送。系统在3到9MHz之间运行,允许在覆盖整个鼠标大脑的视野中,在覆盖整个鼠标大脑的视野中,宽度为130μm,而相同的阵列将用于捕获高分辨率3-D光声数据实时。我们展示了作为体积光声成像能力的系统的体系形成能力,并讨论其潜在的非侵入性监测大脑活动和超声发射的各种影响。

著录项

  • 来源
  • 作者单位

    Univ Zurich Inst Pharmacol & Toxicol CH-8057 Zurich Switzerland|Univ Zurich Fac Med CH-8057 Zurich Switzerland|Univ & ETH Zurich Inst Biomed Engn CH-8093 Zurich Switzerland|Swiss Fed Inst Technol Dept Informat Technol & Elect Engn CH-8093 Zurich Switzerland;

    Univ Zurich Inst Pharmacol & Toxicol CH-8057 Zurich Switzerland|Univ Zurich Fac Med CH-8057 Zurich Switzerland|Univ & ETH Zurich Inst Biomed Engn CH-8093 Zurich Switzerland|Swiss Fed Inst Technol Dept Informat Technol & Elect Engn CH-8093 Zurich Switzerland;

    Univ Zurich Inst Pharmacol & Toxicol CH-8057 Zurich Switzerland|Univ Zurich Fac Med CH-8057 Zurich Switzerland|Univ & ETH Zurich Inst Biomed Engn CH-8093 Zurich Switzerland|Swiss Fed Inst Technol Dept Informat Technol & Elect Engn CH-8093 Zurich Switzerland;

    NYU Langone Hlth Dept Ophthalmol New York NY 10016 USA|NYU Langone Hlth Tech4Hlth Inst New York NY 10016 USA|NYU Langone Hlth Neurosci Inst New York NY 10016 USA;

    Univ Zurich Inst Pharmacol & Toxicol CH-8057 Zurich Switzerland|Univ Zurich Fac Med CH-8057 Zurich Switzerland|Univ & ETH Zurich Inst Biomed Engn CH-8093 Zurich Switzerland|Swiss Fed Inst Technol Dept Informat Technol & Elect Engn CH-8093 Zurich Switzerland;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    Focused ultrasound; optoacoustic imaging; optoacoustic tomography; spherical transducer array; transcranial ultrasound;