A brief survey is given of the calibration methods for hydrophones in the ultrasonic frequency range. The methods presently used in the Physikalisch-Technische Bundesanstalt (PTB), Braunschweig, West Germany, for hydrophone calibrations in the frequency range from 1 to 15 MHz are the primary subject of concern. These methods are the two-transducer reciprocity method for the calibration at discrete frequencies, the time-delay-spectrometry substitution method for quasifrequency continuous calibrations, and the two-transducer reciprocity method with time-delay spectrometry, also for quasi-frequency continuous calibration. Compared with the calibration at discrete frequencies, the expenditure of time for a calibration is considerably reduced in the case of the last-mentioned method. The influencing parameters which affect the evaluation of the measurement uncertainty are briefly discussed for the calibration methods applied at the PTB.
机译:简要调查了超声波频率范围内水听器的校准方法。当前在西德不伦瑞克的物理技术联合会(PTB)中用于1到15 MHz频率范围内水听器校准的方法是主要关注的问题。这些方法是用于离散频率校准的两传感器互易性方法,用于准频率连续校准的时滞谱替代方法,具有时间延迟谱的两传感器互易方法以及准频率连续校准。与离散频率下的校准相比,在最后提到的方法中,校准所花费的时间大大减少了。对于在PTB上使用的校准方法,简要讨论了影响测量不确定度评估的影响参数。