摘要

An online adaptive phased-array ultrasonic imaging system capable of markedly improving the detectability of targets viewed through inhomogeneous media is described. An online adaptive phase correction technique implemented on a research phased-array scanner is described. The theoretical basis for this technique is presented by describing the relationship between the magnitude of phase aberrations and the regional brightness of speckle and pointlike targets. The system currently generates a corrected image in approximately 0.1 s and utilizes no prior knowledge of the aberrating media or the target. The adaptive imaging algorithm uses regional target brightness as a quality factor. The results of in vitro tests with this system using electronic and physical aberrators for both diffuse and pointlike targets are presented.
机译:描述了一种在线自适应相控阵超声成像系统,该系统能够显着提高通过非均匀介质观察到的目标的可检测性。描述了一种在研究相控阵扫描仪上实现的在线自适应相位校正技术。通过描述相差的大小与散斑和点状目标的区域亮度之间的关系,提供了该技术的理论基础。该系统当前在约0.1 s内生成校正图像,并且不使用任何关于像差介质或目标的先验知识。自适应成像算法将区域目标亮度用作质量因子。给出了使用该系统使用电子和物理像差器对弥散和点状目标进行体外测试的结果。