Exposimetry of unfocused pulsed ultrasound

机译:非聚焦脉冲超声的比色法

获取原文
获取原文并翻译 | 示例
           

摘要

Therapeutic applications of unfocused pulsed ultrasound have recently been identified and require specialized exposimetry. Spatial distributions of pressure and several measures of intensity are presented for an unfocused pulsed ultrasound field produced by a PZT-4 transducer with diameter 3.8 cm and center frequency 2 MHz. Distributions of pressure I/sub tp/ and I/sub m/ were adequately uniform over a 4-cm/sup 2/ area in a transverse plane at an axial distance of 7 cm, but not so for I/sub ta/ and I/sub pa/, which both exhibited off-axis asymmetric maxima.
机译:非聚焦脉冲超声的治疗应用近来已经被确定,并且需要专门的辐射测定法。给出了直径为3.8 cm,中心频率为2 MHz的PZT-4换能器产生的未聚焦脉冲超声场的压力空间分布和强度的几种度量。 I / sub tp /和I / sub m /的压力分布在轴向距离为7 cm的横向平面上的4-cm / sup 2 /区域内足够均匀,但对于I / sub ta /和I则不是这样/ sub pa /,它们都表现出离轴不对称最大值。