【24h】

Shock wave pulse pressure after penetration of kidney tissue

机译:肾组织穿透后的冲击波脉压

获取原文
获取原文并翻译 | 示例
           

摘要

Lithotripter generated compressional peak pressure (p/sup +/) was measured after penetrating tissue samples 0.5, 1, 2.5, and 4 cm thick. The samples were taken from fresh pig kidneys and insonated by shock wave pulses with positive peak pressures varying from 14 to 72 MPa. The measurement results allow one to draw some qualitative conclusions concerning the attenuation process in the lithotripter beam and show that in practice the compressional peak pressure values generated in the lithotripter may decrease almost by a factor of two before reaching the kidney stone. On the contrary, theoretical estimates indicate that negative pressures are practically not influenced by attenuation due to their different spectral composition.
机译:在穿透0.5、1、2.5和4 cm厚的组织样本后,测量碎石机产生的压缩峰值压力(p / sup + /)。样品取自新鲜的猪肾脏,并通过正向峰值压力在14至72 MPa之间变化的冲击波脉冲产生声波。测量结果使人们可以得出有关碎石器束衰减过程的一些定性结论,并表明实际上在碎石器中产生的压缩峰值压力值在到达肾结石之前可能几乎减少了两倍。相反,理论估计表明,负压由于其不同的光谱成分而实际上不受衰减影响。