A process for manufacturing epoxy-backed integratednpoly(vinylidene fluoride-trifluoroethylene) [P(VDF-TrFE)] transducersnusing standard microfabrication techniques is reported. The P(VDF-TrFE)nis spin-cast over a silicon membrane, which is backfilled with epoxy tonyield a wide-band, half-wavelength (Λ/2)-resonant device suitablenfor pulse-echo use. The main advantages of the fabrication process arenthe ability to build several devices at a time on a single wafer and thenpotential for integrating electronics on the same substrate with thentransducer. Twelve 31-MHz 2-mm diameter transducers were fabricated on an2-in diameter silicon wafer. A pulse length of about 100 ns was achievednwith a sample device, which exhibited a dynamic range of 40 dB and anninsertion loss of 48 dB. The axial resolution of the transducer innhealthy vascular tissue was estimated to be 79 Μm
机译:报道了一种使用标准的微细加工技术制造环氧支持的集成聚偏二氟乙烯-三氟乙烯[P(VDF-TrFE)]传感器的方法。 P(VDF-TrFE)nis旋铸在硅膜上,硅膜回填了环氧树脂,制成了适用于脉冲回波的宽带,半波长(Λ/ 2)谐振装置。制造过程的主要优点是能够一次在单个晶片上构建多个器件的能力,然后具有利用换能器将电子器件集成在同一基板上的潜力。在直径2英寸的硅晶片上制造了十二个31 MHz直径2毫米的传感器。使用一个采样设备可以实现约100 ns的脉冲长度,该采样设备的动态范围为40 dB,插入损耗为48 dB。换能器对健康的血管组织的轴向分辨率估计为79微米