Estimation of blood velocity with high frequency ultrasound

机译:高频超声估计血流速度

获取原文
获取原文并翻译 | 示例
           

摘要

In order to map blood velocity in small regions near thentransducer, we evaluate the performance of the wideband maximumnlikelihood (WMLE) strategy and infinite impulse response (IIR) filtersnfor blood velocity estimation with a transducer center frequency of 38nMHz. Using a short transmitted pulse and the narrow lateral beam widthnobtained using this frequency, we show that velocities smaller than 1nmm/s can be estimated reliably. In addition, using both changes in thenlocation and magnitude of the peak of the RF correlation, vessels asnsmall as 40 Μm can be visualized in the RF signal and distinguishednfrom stationary tissue. The experimental system also provides thenopportunity to examine changes in flow and in the vessel wall over ancardiac cycle
机译:为了映射靠近传感器的小区域的血流速度,我们评估了宽带最大似然(WMLE)策略和无限冲激响应(IIR)滤波器n在换能器中心频率为38nMHz时进行血流速度估计的性能。使用短发射脉冲和使用该频率获得的窄横向波束宽度,我们表明可以可靠地估计小于1nmm / s的速度。另外,使用RF相关的峰值的位置和大小的变化,可以在RF信号中可视化小至40μm的血管,并将其与静止组织区分开。实验系统还提供了检查心动周期中血流和血管壁变化的机会