Images produced with an intravascular ultrasound system (IVUS) can be distorted because of uncertainty in the instantaneous angular position of a rotating ultrasonic transducer. A rotary encoder placed in proximity to the transducer is required to detect the problem; however, size constraints make a conventional electromechanical or optomechanical encoder difficult to implement. Measurements that test the feasibility of a software-derived encoder, based of the rate of decorrelation of ultrasonic RF lines with angle, are reported. Provided that the instantaneous angular velocity of the transducer can be measured, adjustments can be made to the pulse rate of the transducer, which would eliminate the image distortion.
机译:由于旋转超声换能器的瞬时角度位置不确定,使用血管内超声系统(IVUS)产生的图像可能会失真。需要在传感器附近放置一个旋转编码器以检测问题。然而,尺寸限制使得传统的机电编码器难以实现。报告了基于超声RF线与角度的去相关速率测试软件衍生编码器可行性的测量。只要可以测量换能器的瞬时角速度,就可以对换能器的脉冲频率进行调整,这将消除图像失真。