The performance of ultrasono-therapy transducer with contact detection by using the impedance phase change is described. Usually a therapy transducer is designed with a /spl lambda//2 frontal plate glued to a PZT-4 piezoceramic. This plate ensures a good mechanical protection of the piezoceramic with a corresponding high-transmission energy. Normally this transducer is operated at the minimum at the frequency of the impedance module of its input electric impedance, but this operation point is affected by the shift caused by the expected temperature increase. This shift could be higher than the narrow bandwidth presented. As a result we obtain a decrease in the power level for medical treatment. Usually electronic drivers with automatic control that follow the frequency change are designed, however, the relatively narrow bandwidth introduces difficulty in the design. Another frequency operation point is presented here and analyzed using the criteria of the maximum of the impedance phase with a wider bandwidth than in the previous case. Simulation with mechanical losses are presented with experimental results that show the convenience of this criteria for practical application.
机译:描述了通过使用阻抗相变进行接触检测的超声波治疗换能器的性能。通常,治疗换能器的设计是将/ spl lambda // 2前置板粘贴到PZT-4压电陶瓷上。该板通过相应的高传输能量确保了对压电陶瓷的良好机械保护。通常,此换能器以其输入电阻抗的阻抗模块的频率最小运行,但此工作点受预期温度升高引起的偏移的影响。此偏移可能高于显示的窄带宽。结果,我们降低了医疗的功率水平。通常设计具有跟随频率变化的自动控制的电子驱动器,但是,相对窄的带宽在设计中引入了困难。此处介绍了另一个频率工作点,并使用阻抗相的最大值(比以前的情况下具有更大的带宽)的标准进行了分析。带有机械损耗的仿真与实验结果一起展示,这些结果表明了该标准在实际应用中的便利性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号