Comparative evaluation of despeckle filtering in ultrasound imaging of the carotid artery

机译:去斑滤波在颈动脉超声成像中的比较评估

获取原文
获取原文并翻译 | 示例
           

摘要

It is well-known that speckle is a multiplicative noise that degrades the visual evaluation in ultrasound imaging. The recent advancements in ultrasound instrumentation and portable ultrasound devices necessitate the need of more robust despeckling techniques for enhanced ultrasound medical imaging for both routine clinical practice and teleconsultation. The objective of this work was to carry out a comparative evaluation of despeckle filtering based on texture analysis, image quality evaluation metrics, and visual evaluation by medical experts in the assessment of 440 (220 asymptomatic and 220 symptomatic) ultrasound images of the carotid artery bifurcation. In this paper a total of 10 despeckle filters were evaluated based on local statistics, median filtering, pixel homogeneity, geometric filtering, homomorphic filtering, anisotropic diffusion, nonlinear coherence diffusion, and wavelet filtering. The results of this study suggest that the first order statistics filter lsmv, gave the best performance, followed by the geometric filter gf4d, and the homogeneous mask area filter lsminsc. These filters improved the class separation between the asymptomatic and the symptomatic classes based on the statistics of the extracted texture features, gave only a marginal improvement in the classification success rate, and improved the visual assessment carried out by the two experts. More specifically, filters lsmv or gf4d can be used for despeckling asymptomatic images in which the expert is interested mainly in the plaque composition and texture analysis; and filters lsmv, gf4d, or lsminsc can be used for the despeckling of symptomatic images in which the expert is interested in identifying the degree of stenosis and the plaque borders. The proper selection of a despeckle filter is very important in the enhancement of ultrasonic imaging of the carotid artery. Further work is needed to evaluate at a larger scale and in clinical practice the performance of the proposed despeckle filters in the automated segmentation, texture analysis, and classification of carotid ultrasound imaging.
机译:众所周知,斑点是一种乘性噪声,会降低超声成像的视觉评价。超声仪器和便携式超声设备的最新发展需要对常规临床实践和远程咨询都需要更强大的去斑点技术,以增强超声医学成像。这项工作的目的是基于纹理分析,图像质量评估指标和医学专家的视觉评估,对颈动脉分叉的440张(220例无症状和220例有症状)超声图像进行比较评估。 。本文基于局部统计,中值滤波,像素均匀性,几何滤波,同态滤波,各向异性扩散,非线性相干扩散和小波滤波,总共评估了10个散斑滤波器。这项研究的结果表明,一阶统计滤波器lsmv表现最佳,其次是几何滤波器gf4d和齐次掩膜面积滤波器lsminsc。这些过滤器基于提取的纹理特征的统计数据,改善了无症状和有症状类别之间的类别分离,仅略微提高了分类成功率,并改善了两位专家进行的视觉评估。更具体地说,滤镜lsmv或gf4d可用于使无症状图像散斑,在这些无症状图像中,专家主要关注菌斑的组成和质地分析;过滤器lsmv,gf4d或lsminsc可以用于症状图像的去斑处理,在这些图像中,专家对确定狭窄程度和斑块边界感兴趣。去斑滤波器的正确选择对于增强颈动脉的超声成像非常重要。需要做更多的工作来评估更大的规模,并在临床实践中,建议的去斑点滤波器在颈动脉超声成像的自动分割,纹理分析和分类中的性能。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号