Single-pulse tissue doppler using synthetic transmit beams

机译:使用合成发射束的单脉冲组织多普勒

获取原文
获取原文并翻译 | 示例
           

摘要

Tissue Doppler imaging (TDI) is a common technique for investigating myocardial function. Typically, B-mode data and TDI data are recorded using separate acquisitions and combined into a single, color overlaid image. In this work, we present a novel method for TDI imaging, where both TDI and B-mode are created from the same acquisition. Velocities are calculated from the phase shift between neighboring transmit events in the B-mode scan; hence the name singlepulse tissue Doppler (SPTD). Using a novel transmit beam interleaving pattern, this method provides TDI and B-mode at the same high frame rate with an adjustable Nyquist velocity limit. Through simulations and measurements, this work investigates the bias and variance of the SPTD velocities and compares the estimates to those of the conventional TDI autocorrelation estimation method. The results showed that the method introduces an additional bias and variance in the velocity estimates compared with conventional TDI. However, by applying bias compensation, the SPTD velocity estimates were close to those of regular TDI. Using SPTD, the whole left ventricle was imaged within a 65-degree sector at a frame rate of 110 frames per second (43 transmissions per frame).
机译:组织多普勒成像(TDI)是研究心肌功能的常用技术。通常,使用单独的采集记录B模式数据和TDI数据,并将其合并为一个彩色叠加图像。在这项工作中,我们提出了一种用于TDI成像的新颖方法,其中TDI和B模式都是从同一采集中创建的。根据B模式扫描中相邻发射事件之间的相移来计算速度;因此,名称为单脉冲组织多普勒(SPTD)。使用一种新颖的发射波束交织模式,该方法以相同的高帧速率和可调节的奈奎斯特速度极限提供TDI和B模式。通过模拟和测量,这项工作研究了SPTD速度的偏差和方差,并将估计值与常规TDI自相关估计方法的估计值进行了比较。结果表明,与传统的TDI相比,该方法在速度估计中引入了额外的偏差和方差。但是,通过应用偏差补偿,SPTD速度估算值接近常规TDI。使用SPTD,整个左心室在65度扇区内以每秒110帧的帧速率(每帧43次传输)成像。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号