首页> 外文期刊>IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control >Absolute ultrasound perfusion parameter quantification of a tissue-mimicking phantom using bolus tracking [Correspondence]
【24h】

Absolute ultrasound perfusion parameter quantification of a tissue-mimicking phantom using bolus tracking [Correspondence]

机译:使用推注跟踪对模拟组织幻象的绝对超声灌注参数进行量化[通讯]

获取原文
获取原文并翻译 | 示例
           

摘要

This study presents three methods for absolute quantification in ultrasound perfusion analysis based on bolus tracking. The first two methods deconvolve the perfusion time sequence with a measured AIF, using a nonparametric or a parametric model of the tissue residue function, respectively. The third method is a simplified approach avoiding deconvolution by assuming a narrow AIF. A phantom with a dialyzer filter as a tissue-mimicking model was used for evaluation. Estimated mean transit times and blood volumes were compared with the theoretical values. A match with a maximum error of 12% was achieved.
机译:这项研究提出了三种基于推注跟踪的超声灌注分析绝对定量方法。前两种方法分别使用组织残差函数的非参数模型或参数模型对灌注时间序列与测量的AIF进行卷积。第三种方法是一种简化方法,通过假设AIF较窄来避免反卷积。具有透析器滤光片的假体作为组织模仿模型用于评估。将估计的平均通过时间和血容量与理论值进行比较。实现了最大误差为12%的匹配。