首页> 外文期刊>IEEE transactions on biomedical circuits and systems >A Two-Electrode, Double-Pulsed Sensor Readout Circuit for Cardiac Troponin I Measurement
【24h】

A Two-Electrode, Double-Pulsed Sensor Readout Circuit for Cardiac Troponin I Measurement

机译:用于心肌肌钙蛋白的双电极,双脉冲传感器读数电路测量

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

This paper presents a pulse-stimulus sensor readout circuit for use in cardiovascular disease examinations. The sensor is based on a gold nanoparticle plate with an antibody post-modification. The proposed system utilizes gated pulses to detect the biomarker Cardiac Troponin I in an ionic solution. The characteristic of the electrostatic double-layer capacitor generated by the analyte is related to the concentration of Cardiac Troponin I in the solvent. After sensing by the transistor, a current-to-frequency converter (I-to-F) and delay-line-based time-to-digital converter (TDC) convert the information into a series of digital codes for further analysis. The design is fabricated in a 0.18-μm standard CMOS process. The chip occupies an area of 0.92 mm 2 and consumes 125 μW. In the measurements, the proposed circuit achieved a 1.77 Hz/pg-mL sensitivity and 72.43 dB dynamic range.
机译:本文介绍了用于心血管疾病检查的脉冲刺激传感器读数电路。传感器基于具有抗体后的金纳米颗粒板。所提出的系统利用门控脉冲在离子溶液中检测生物标志物心肌肌钙蛋白I。由分析物产生的静电双层电容器的特性与溶剂中心肌肌钙蛋白I的浓度有关。在晶体管感测后,电流转换器(I-TO-F)和基于延时线的时间转换器(TDC)将信息转换为一系列数字代码以进行进一步分析。该设计采用0.18-μm标准CMOS工艺制造。该芯片占据0.92 mm 2 ,消耗125μw。在测量中,所提出的电路达到1.77Hz / pg-ml灵敏度和72.43dB动态范围。