...
首页> 外文期刊>Environmental Monitoring and Assessment >INTER-LABORATORY COMPARISON OF NO_2 AND SO_2 GENERATED BY DYNAMIC DILUTION SYSTEM UNDER LABORATORY CONDITIONS: A TECHNICAL DISCUSSION
【24h】

INTER-LABORATORY COMPARISON OF NO_2 AND SO_2 GENERATED BY DYNAMIC DILUTION SYSTEM UNDER LABORATORY CONDITIONS: A TECHNICAL DISCUSSION

机译:实验室条件下动态稀释系统产生的NO_2和SO_2的实验室间比较:技术探讨

获取原文
获取原文并翻译 | 示例
           

摘要

A workshop on analytical quality control (AQC) of ambient air quality measurement methods for nitrogen dioxide (NO_2) and sulphur dioxide (SO_2) was conducted by Central Pollution Control Board (CPCB) for officials involved in National Ambient Air Quality Monitoring (NAAQM) in India. Concentrations of NO_2 and SO_2 were generated by dynamic dilution system under laboratory conditions at low and high levels and measured using static dilution system and wet chemical methods laid down by CPCB under section 16(2)(h) of the air act 1981. CPCB provided the measured values as reference values for comparing the means obtained by the officials participated from thirteen organizations. A tolerance limit of ± 15% of the reference values was specified to accept the results. Generated concentrations, which were unknown to the participants, were measured using gaseous sampling assembly (Envirotech APM 411, New Delhi, India), and wet chemical methods laid down by CPCB i.e. the same methodology which is used by the organizations to generate the data of NO_2 and SO_2 in ambient air. Simultaneously, concentrations were checked by CPCB using automatic analyzers as a check on reference concentration. It is observed that results of automatic analyzers for NO_2 and SO_2 were within a tolerance of ± 5% with %RSD below 3. On the other hand, results of most of the participants showed variability in the measurements with %RSD ranging between ± 0.8 and ± 88.6 and exceedences of means from the tolerance limit with bias ranging between 1.4 and -59%. To check the cause of high variability in the measurements obtained under identical conditions, duplicate sampling was performed by one of the participants for SO_2 at low concentration level. In this study, results of wet chemical methods, automatic analyzers and results of duplicate sampling are analysed statistically to assess the cause of high variability in the measurements. Analysis of t-test and analysis of variance (ANOVA) showed highly significant results for NO_2 and SO_2 at high concentration levels (α 0.05) and for SO_2 at both the levels (α 0.01) respectively indicating some bias is existing either in the sampling or in analytical technique. Duplicate sampling performed to check precision in parallel measurements showed high %RSD indicating the presence of systematic error in sampling technique as the same calibration factor (CF) was used to measure the concentration of duplicate samples. Statistical analysis of flow rates of duplicate sampling showed that the sampling assembly could not maintain the constant flow rate within the ± 10% with that measured at the start of the sampling. This resulted in high %RSD and deviation from the reference values for the results of most of the participants, even after accepting ± 15% tolerance limit. There is a need to improve and evaluate this gaseous sample collection device under laboratory conditions to generate reliable database of NO_2 and SO_2 in ambient air.
机译:中央污染控制委员会(CPCB)为参与全国环境空气质量监测(NAAQM)的官员举办了关于二氧化氮(NO_2)和二氧化硫(SO_2)的环境空气质量测量方法的分析质量控制(AQC)研讨会。印度。 NO_2和SO_2的浓度是通过动态稀释系统在实验室条件下在低水平和高水平下产生的,并使用静态稀释系统和CPCB根据1981年《空气法》第16(2)(h)条规定的湿化学方法进行测量。测量值作为参考值,用于比较13个组织的官员参加的调查所得的平均值。指定公差范围为参考值的±15%以接受结果。参与者不知道的产生浓度是使用气体采样组件(Envirotech APM 411,印度新德里)和CPCB规定的湿化学方法测量的,即组织使用的相同方法来生成数据。环境空气中的NO_2和SO_2。同时,使用自动分析仪通过CPCB检查浓度,以检查参考浓度。可以观察到,NO_2和SO_2的自动分析仪的结果在±5%的容差范围内,%RSD低于3。另一方面,大多数参与者的结果显示,%RSD的测量范围在±0.8至0.5之间。 ±88.6,超出公差范围的均值,偏差在1.4到-59%之间。为了检查在相同条件下获得的测量结果存在高变异性的原因,参与者之一对低浓度的SO_2进行了重复采样。在这项研究中,对湿化学方法,自动分析仪的结果和重复采样的结果进行了统计分析,以评估测量结果高度可变的原因。 t检验分析和方差分析(ANOVA)表明,高浓度水平(α0.05)的NO_2和SO_2和两个浓度水平(α0.01)的SO_2均具有非常显着的结果,表明在采样或采样中存在一定偏差。在分析技术上。为检查平行测量的准确性而进行的重复采样显示出较高的%RSD,表明在采样技术中存在系统误差,因为使用相同的校准因子(CF)来测量重复样品的浓度。对重复采样的流量进行统计分析表明,采样组件无法将恒定流量保持在采样开始时的±10%以内。即使接受了±15%的公差极限,这也导致了较高的%RSD和大多数参与者的结果与参考值的偏差。需要在实验室条件下改进和评估这种气态样品收集装置,以生成环境空气中NO_2和SO_2的可靠数据库。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号