Factors Affecting Repeatability of Foveal Avascular Zone Measurement Using Optical Coherence Tomography Angiography in Pathologic Eyes

机译:在病理眼中使用光学相干断层造影血管造影影响心肺缺血区测量的可重复性的因素

获取原文
           

摘要

Purpose: To determine factors that may affect the repeatability of the foveal avascular zone (FAZ) measurement obtained using optical coherence tomography angiography (OCTA) including instrument type, image segmentation, image quality, and fundus pathology. Patients and Methods: This prospective single-center study enrolled 43 subjects (85 eyes) with retinal vasculopathy, macular edema, optic pathology or normal contralateral eye. The macula was imaged twice using Optovue Angiovue and once using Cirrus Angioplex to obtain 3x3mm OCTA images centered on the fovea. Images were?generated by the same operator within 30?mins. The FAZ size for the entire retinal thickness (“overall FAZ”) was measured automatically using the OCTA software. The FAZ size of the superficial and deep retinal vascular plexus layers was measured manually using the enface OCTA images of the segmented layers and Image J analysis. Intraclass correlations coefficient (ICC) was calculated to determine repeatability. Results: For the overall FAZ measurement, repeatability was excellent (ICC 0.953 right eye, 0.938, left eye) using the same machine (intra-instrument) and somewhat lower but still good?to?excellent (ICC 0.803 right eye, 0.917 left eye) using machines made by different vendors (inter-instrument). For the segmented layers, intra-instrument repeatability of FAZ measurement was excellent (ICC 0.95) for both plexus layers. Inter-instrument repeatability was good for the superficial plexus layer (ICC 0.86 right eye, 0.88 left eye) but reduced for the deep plexus layer (ICC 0.63 right eye, 0.57 left eye). Suboptimal image quality and presence of retinal vasculopathy and macular edema tended to reduce FAZ repeatability but?to a lesser degree. Conclusion: Inter- and intra-instrument repeatability of the overall FAZ measurement was high using commercial OCTA instruments and only mildly reduced by suboptimal image quality and fundus pathology. For segmented layers, intra-instrument repeatability remained high but inter-instrument repeatability was reduced for the deep plexus layer.
机译:目的:确定可能影响使用光学相干断层造影血管造影(OctA)获得的可变血管区(FAZ)测量的可重复性的因素,包括仪器类型,图像分割,图像质量和眼底病理。患者和方法:该前瞻性单中心研究注册了43名受试者(85只眼),具有视网膜血管病,黄斑水肿,视神经理或正常对侧眼。使用Optiovue Angiovue拍摄两次蛋白孢子,并使用Cirrus angioplex获得3x3mm octa图像以Fovea为中心。图像是?由30次在30?min的相同操作员生成。使用Octa软件自动测量整个视网膜厚度(“整体FAZ”)的FAZ尺寸。使用分段层的enface Octa图像和图像J分析手动测量浅表和深视网膜血管孔隙层的Faz尺寸。计算脑内相关系数(ICC)以确定可重复性。结果:对于整体FAZ测量,可重复性优异(ICC 0.953右眼,0.938,左眼)使用同一台机器(仪器仪器)和略低但仍然好?到?优秀(ICC 0.803右眼,0.917左眼0.917 )使用不同供应商(仪器间)制造的机器。对于分段层,对于丛层,FAZ测量的仪器内可重复性优异(ICC> 0.95)。仪器间重复性适用于浅表丛层(ICC 0.86右眼,左眼0.88),但为深层丛层(ICC 0.63右眼,0.57左眼)减少。视网膜血管病变和黄斑水肿的次优图像质量和存在倾向于降低FAZ可重复性,但呢?到较小程度。结论:使用商业Octa Instruments的整体FAZ测量的仪器间可重复性高,并且仅通过次优图像质量和眼底病理略微降低。对于分段层,仪器内可重复性仍然很高,但深层丛层的仪器间可重复性降低。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号