首页> 外文期刊>Journal of Clinical Microbiology >Comparison of BD Phoenix, Vitek 2, and MicroScan Automated Systems for Detection and Inference of Mechanisms Responsible for Carbapenem Resistance in Enterobacteriaceae
【24h】

Comparison of BD Phoenix, Vitek 2, and MicroScan Automated Systems for Detection and Inference of Mechanisms Responsible for Carbapenem Resistance in Enterobacteriaceae

机译:BD Phoenix,Vitek 2和MicroScan自动化系统在肠杆菌科中对碳青霉烯耐药性负责机制的检测和推断的比较

获取原文
           

摘要

We assessed the ability of three commercial systems to infer carbapenem resistance mechanisms in 39 carbapenemase-producing and 16 other carbapenem-resistant Enterobacteriaceae. The sensitivity/specificity values for “flagging” a likely carbapenemase were 100%/0% (BD Phoenix), 82 to 85%/6 to 19% (MicroScan), and 74%/38% (Vitek 2), respectively. OXA-48 producers were poorly detected, but all systems reliably detected isolates with KPC and most with metallo-carbapenemases.
机译:我们评估了三种商业系统推断39种碳青霉烯酶和16种其他对碳青霉烯耐药的 Enterobacteriaceae 的碳青霉烯耐药机制的能力。 “标记”可能的碳青霉烯酶的敏感性/特异性值分别为100%/ 0%(BD Phoenix),82%至85%/ 6%至19%(MicroScan)和74%/ 38%(Vitek 2)。 OXA-48生产者的检测较差,但所有系统均可靠地检测到含有KPC的分离株,大部分含有金属-卡宾烯酶。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号