This study compared the Biomic automated well reader results to the MicroScan WalkAway results for reading MicroScan antimicrobial susceptibility and identification panels at four different sites. Routine fresh clinical isolates and quality control (QC) organisms were tested at each study site. A total of 46,176 MicroScan panel drug-organism combinations were read. The Biomic category agreement for 3,117 Gram-negative bacteria was 98.4%, with 1.4% minor and 0.2% major discrepancies. The Biomic category agreement for 5,233 Gram-positive bacteria was 98.7%, with 0.9% minor, 0.3% major, and 0.1% very major errors. Essential agreement, defined as Biomic results that were within ±1 2-fold dilution of the MicroScan results, was 99.3% for Gram-negative bacteria and 98.3% for Gram-positive bacteria. Biomic reading of MicroScan identification panels provided an overall agreement (first- and second-choice organism match) of 99.5% with 846 Gram-negative isolates and 99.5% with 430 Gram-positive isolates. These results suggest that the Biomic automated reader can provide accurate reading of MicroScan panels and has the capability of a visual panel read for manual adjustment of results.
机译:这项研究将Biomic自动孔读取器结果与MicroScan WalkAway结果进行了比较,以在四个不同的位置读取MicroScan抗菌药敏性和鉴定标本。在每个研究地点测试了常规的新鲜临床分离株和质量控制(QC)生物。总共读取了46,176种MicroScan面板药物生物组合。 3,117例革兰氏阴性细菌的生物组分类协议为98.4%,次要差异为1.4%,主要差异为0.2%。 5,233例革兰氏阳性细菌的Biomic类别协议为98.7%,其中小错误为0.9%,大错误为0.3%,非常重大错误为0.1%。基本一致性定义为在MicroScan结果的±1 2倍稀释范围内的Biomic结果,对于革兰氏阴性细菌为99.3%,对于革兰氏阳性细菌为98.3%。 MicroScan识别面板的生物读物显示,与846株革兰氏阴性菌分离出的细菌(99.5%)和与430革兰氏阳性菌分离出的细菌(99.5%)的总体一致性。这些结果表明,Biomic自动读取器可以准确读取MicroScan面板,并具有可视面板读取功能,可以手动调整结果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号