首页> 外文期刊>Journal of Clinical Microbiology >Erroneous Reporting of Coagulase-Negative Staphylococci as Kocuria spp. by the Vitek 2 System
【24h】

Erroneous Reporting of Coagulase-Negative Staphylococci as Kocuria spp. by the Vitek 2 System

机译:凝血酶阴性葡萄球菌为科氏菌属的错误报道。通过Vitek 2系统

获取原文
   

获取外文期刊封面封底 >>

       

摘要

Misidentification of coagulase-negative staphylococci (CoNS) may delay appropriate treatment. We investigated 20 clinical isolates identified as Kocuria spp. by the Vitek 2 system. All were identified as CoNS by 16S rRNA gene sequencing (18 Staphylococcus epidermidis, 1 Staphylococcus haemolyticus). Four Kocuria isolates were shown to be identical to CoNS from the same patient by pulsed-field gel electrophoresis. Isolates identified by Vitek 2 as Kocuria most likely represent misidentified CoNS, and if clinically indicated, should be investigated further by genomic methods.
机译:凝固酶阴性葡萄球菌(CoNS)的错误识别可能会延迟适当的治疗。我们调查了20个临床分离株,鉴定为 Kocuria spp。通过Vitek 2系统。通过16S rRNA基因测序(表皮葡萄球菌,1 溶血葡萄球菌)将其全部确定为CoNS。通过脉冲场凝胶电泳显示,来自同一患者的四个 Kocuria 分离株与CoNS相同。 Vitek 2鉴定为 Kocuria 的分离株很可能代表了错误鉴定的CoNS,如果临床上有指示,应通过基因组方法进一步研究。