Nucleic acid amplification tests such as the BDProbeTec ET (BDPT) system are more prone to reproducibility problems than are antigen detection tests and culture. A repeat testing algorithm for all samples with method other than acceleration (MOTA) scores greater than or equal to the cutoff value (2,000) was developed for the BDPT system and applied in a clinical laboratory setting. All positive samples were retested, and if the result of the second test was below the cutoff value, a third test was performed to resolve the discrepancy. Overall, 11 (5.3%) of 207 samples initially positive for Chlamydia trachomatis and 11 (10.7%) of 103 samples initially positive for Neisseria gonorrhoeae were not confirmed by repeat testing of the original sample. Poor reproducibility was associated with low-positive MOTA scores (2,000 to 9,999) for both analytes. Only 21 (80.8%) of 26 low-positive samples in the C. trachomatis test and 4 (33.3%) of 12 low-positive samples in the N. gonorrhoeae test retested as positive. The reproducibility of both tests with samples with initial MOTA scores of ≥10,000 increased to 96.7%. The data suggest that retesting of low-positive samples is warranted and could reduce the number of potentially false-positive test results.
机译:诸如BDProbeTec ET(BDPT)系统之类的核酸扩增测试比抗原检测测试和培养更容易出现再现性问题。针对BDPT系统开发了针对所有样品的重复测试算法,该方法的加速度(MOTA)得分均大于或等于临界值(2,000),而该方法已应用于BDPT系统,并应用于临床实验室环境。重新测试所有阳性样品,如果第二次测试的结果低于临界值,则进行第三次测试以解决差异。总体而言,通过反复测试未确认沙眼衣原体为阳性的207个样本中的11个(5.3%)和最初为淋病奈瑟氏菌呈阳性的103个样品中的11个(10.7%)。原始样本。两种分析物的重现性差均与MOTA分数较低(2,000至9,999)有关。在 C的26个低阳性样本中,只有21个(80.8%)。沙眼测试和 N中的12个低阳性样本中有4个(33.3%)。淋病测试再次呈阳性。初始MOTA得分≥10,000的样品的两种测试的重现性均提高到96.7%。数据表明,对低阳性样品进行重新测试是有必要的,并且可以减少潜在的假阳性测试结果的数量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号