Evaluation of the Fully Automated BACTEC MGIT 960 System for Testing Susceptibility of Mycobacterium tuberculosis to Pyrazinamide, Streptomycin, Isoniazid, Rifampin, and Ethambutol and Comparison with the Radiometric BACTEC 460TB Method

机译:评估用于测试结核分枝杆菌对吡嗪酰胺,链霉素,异烟肼,利福平和乙胺丁醇敏感性的全自动BACTEC MGIT 960系统的评估,并与辐射BACTEC 460TB方法进行比较

获取原文
           

摘要

The performance of the fully automated BACTEC MGIT 960 (M960) system for the testing of Mycobacterium tuberculosis susceptibility to streptomycin (SM), isoniazid (INH), rifampin (RMP), ethambutol (EMB), and pyrazinamide (PZA) was evaluated with 100 clinical isolates and compared to that of the radiometric BACTEC 460TB (B460) system. The agar proportion method and the B460 system were used as reference methods to resolve the discordant results for SM, INH, RMP, and EMB (a combination known as SIRE) and PZA, respectively. The overall agreements were 96.3% for SIRE and 92% for PZA. For SIRE, a total of 26 discrepancies were found and were resolved in favor of the M960 system in 8 cases and in favor of the B460 system in 18 cases. The M960 system produced 8 very major errors (VME) and 10 major errors (ME), while the B460 system showed 4 VME and 4 ME. No statistically significant differences were found. Both systems exhibited excellent performance, but a higher number of VME was observed with the M960 system at the critical concentrations of EMB and SM. For PZA, a total of eight discrepancies were observed and were resolved in favor of the M960 system in one case and in favor of the B460 system in seven cases; no statistically significant differences were found. The M960 system showed four VME and three ME. The mean times to report overall PZA results and resistant results were 8.2 and 9.8 days, respectively, for the M960 system and 7.4 and 8.1 days, respectively, for the B460 system. Statistically significant differences were found. The mean times to report SIRE results were 8.3 days for the M960 system and 8.2 days for the B460 system. No statistically significant differences were found. Twelve strains tested for SIRE susceptibility and seven strains tested for PZA susceptibility had been reprocessed because of contamination. In conclusion, the M960 system can represent a valid alternative to the B460 for M. tuberculosis susceptibility testing; however, the frequent contamination of the tests needs to be improved.
机译:全自动BACTEC MGIT 960(M960)系统用于测试结核分枝杆菌(SM),异烟肼(INH),利福平(RMP),乙胺丁醇(EMB)和吡嗪酰胺的易感性的性能(PZA)已用100种临床分离株进行了评估,并与辐射BACTEC 460TB(B460)系统进行了比较。使用琼脂比例法和B460系统作为参考方法,分别解决SM,INH,RMP和EMB(称为SIRE的组合)和PZA的不一致结果。 SIRE和PZA的总体协议分别为96.3%和92%。对于SIRE,总共发现了26个差异,并解决了8例中使用M960系统和18例中B460系统的差异。 M960系统产生了8个非常严重的错误(VME)和10个重大错误(ME),而B460系统显示了4个VME和4个ME。没有发现统计学上的显着差异。两种系统均表现出出色的性能,但是在临界浓度EMB和SM下,使用M960系统观察到了更高数量的VME。对于PZA,总共观察到8个差异,其中有1个案例支持M960系统,有7个案例支持B460系统。没有发现统计学上的显着差异。 M960系统显示四个VME和三个ME。对于M960系统,报告总体PZA结果和耐药结果的平均时间分别为8.2天和9.8天,对于B460系统,分别为7.4和8.1天。发现统计学上的显着差异。报告SIRE结果的平均时间在M960系统中为8.3天,在B460系统中为8.2天。没有发现统计学上的显着差异。由于污染,对十二种测试过的SIRE敏感性菌株和七种测试过的PZA敏感性菌株进行了重新处理。总之,对于 M,M960系统可以代表B460的有效替代品。结核病敏感性测试;但是,测试的频繁污染需要改善。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号