Evaluation of an Automated Sample Preparation Protocol for Quantitative Detection of Hepatitis C Virus RNA

机译:定量检测丙型肝炎病毒RNA的自动样品制备方案的评估

获取原文
           

摘要

The COBAS AMPLIPREP instrument for automated sample preparation has recently been introduced. In this study, the COBAS AMPLIPREP/COBAS AMPLICOR HCV MONITOR test, which includes this new molecular device, was evaluated and compared to the COBAS AMPLICOR HCV MONITOR test, which includes a manual extraction protocol. Interassay and intra-assay variation, precision, and linearity were determined, and a total of 130 clinical specimens were investigated. For determination of interassay variation, coefficients of variation were found to be between 9 and 59% for the COBAS AMPLIPREP/COBAS AMPLICOR HCV MONITOR test and between 13 and 69% for the COBAS AMPLICOR HCV MONITOR test. For determination of intra-assay variation, coefficients of variation were found to be between 7 and 13% for the COBAS AMPLIPREP/COBAS AMPLICOR HCV MONITOR test and between 8 and 16% for the COBAS AMPLICOR HCV MONITOR test. When precision of the COBAS AMPLIPREP/COBAS AMPLICOR HCV MONITOR test was tested, all results were found to be within ±0.5 log of the expected results. Determination of linearity resulted in a quasilinear curve over 3 logs. When clinical samples were tested with the COBAS AMPLIPREP/COBAS AMPLICOR HCV MONITOR test and compared with the COBAS AMPLICOR HCV MONITOR test, all results were found within ±0.5 log. In conclusion, the assay, which included the new molecular device, proved to be suitable for the routine molecular laboratory. It was found to be laborsaving and easy to handle.
机译:最近推出了用于自动样品制备的COBAS AMPLIPREP仪器。在这项研究中,评估了包括这种新分子装置的COBAS AMPLIPREP / COBAS AMPLICOR HCV MONITOR测试,并将其与包括手动提取方案的COBAS AMPLICOR HCV MONITOR测试进行了比较。测定了批间和批内变异,准确性和线性,并调查了130个临床标本。为了确定批间差异,对于COBAS AMPLIPREP / COBAS AMPLICOR HCV MONITOR测试,变异系数介于9%至59%之间;对于COBAS AMPLICOR HCV MONITOR测试,变异系数介于13%至69%之间。为了确定批内变异,对于COBAS AMPLIPREP / COBAS AMPLICOR HCV MONITOR测试,变异系数为7%至13%;对于COBAS AMPLICOR HCV MONITOR测试,变异系数为8%至16%。对COBAS AMPLIPREP / COBAS AMPLICOR HCV MONITOR测试的精度进行测试时,发现所有结果均在预期结果的±0.5 log之内。线性的确定导致3个对数上的准线性曲线。当用COBAS AMPLIPREP / COBAS AMPLICOR HCV MONITOR测试对临床样品进行测试并与COBAS AMPLICOR HCV MONITOR测试进行比较时,发现所有结果均在±0.5 log之内。总之,包含新分子装置的测定方法被证明适用于常规分子实验室。发现它省力并且易于处理。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号