Quantitative cultures of bronchoalveolar lavage (BAL) fluid are important in the diagnosis of ventilator-associated pneumonia, and calibrated loops are commonly used to set up these cultures. In this study, the performances of calibrated 0.010- and 0.001-ml loops in the transfer of BAL fluid were determined. Five loops of one lot from seven manufacturers were tested. Calibrations were performed by the gravimetric method (0.010-ml loops) and the colorimetric method (0.001-ml loops). Most of the 0.010-ml loops displayed a precision that was less than 10%, but six of them showed very poor accuracies as they transferred a deficiency (nichrome loops) or an excess (disposable loops) of BAL fluid that exceeded ±10%. The mean maximum and minimum BAL fluid volumes delivered by the 0.010-ml loops differed by a factor 3. The 0.001-ml loops displayed acceptable precision. Five of them showed inaccuracies of ≤±10%, and mean maximum and minimum BAL fluid volumes had a range of a factor of 2. For all loops, the volumes of BAL fluid sampled were larger than the volumes of reagent-grade water sampled. Results of the colony counting experiments confirmed these findings and revealed a high intra-assay variability for the 0.001-ml loops. We conclude that, when BAL fluid samples are cultured with calibrated loops, (i) proper verification of the calibration of these loops is mandatory, (ii) calibrations should be performed with BAL fluid as the test solution, and (iii) borderline quantitative culture results should be interpreted with knowledge of the inaccuracy values of these loops.
机译:支气管肺泡灌洗液(BAL)的定量培养在呼吸机相关性肺炎的诊断中很重要,通常使用校准环建立这些培养物。在这项研究中,确定了在BAL液的转移中校准的0.010-ml和0.001-ml定量环的性能。测试了来自七个制造商的一批的五个回路。通过重量法(0.010 ml定量环)和比色法(0.001 ml定量环)进行校准。大部分0.010 ml的定量环显示的准确度均低于10%,但其中有六个却显示出非常差的准确度,因为它们转移了BAL液的不足(镍铬合金环)或过量(一次性环)超过±10%。 0.010 ml定量环输送的平均最大和最小BAL液量相差3倍。0.001 ml定量环显示可接受的精度。其中五个显示不准确度≤±10%,并且平均最大和最小BAL流体体积的范围为2倍。对于所有回路,采样的BAL流体体积均大于试剂级水的采样体积。菌落计数实验的结果证实了这些发现,并揭示了0.001 ml定量环的批内分析差异很大。我们得出的结论是,当用校准的定量环培养BAL流体样品时,(i)必须对这些定量环的校准进行正确验证,(ii)应使用BAL流体作为测试溶液进行校准,并且(iii)边界定量培养应该在知道这些循环的不准确性值的情况下解释结果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号