Stabilities of free and complexed human immunodeficiency virus p24 antigens during short- and long-term storage.

机译:游离和复杂的人类免疫缺陷病毒p24抗原在短期和长期储存过程中的稳定性。

获取原文
           

摘要

By the standard p24 assay there was a 25 to 27% decrease in free p24 antigen in serum after storage at 4 degrees C over 14 days but no loss at -70 degrees C. There was no loss at either temperature by the immune complex dissociation (ICD) procedure. Furthermore, there was no significant loss of detectable p24 in serum by either the ICD or the standard p24 assay after 700 days of storage at -70 degrees C.
机译:通过标准p24分析,在4°C下储存14天后,血清中的游离p24抗原减少了25%至27%,但在-70°C下没有损失。免疫复合物的解离在任一温度下均没有损失( ICD)程序。此外,在-70摄氏度下保存700天后,通过ICD或标准p24测定,血清中可检测到的p24均无明显损失。