Effect of Errors in the Sequence of Optical Densities from the Roche AMPLICOR HIV-1 MONITOR Assay on the Validity of Assay Results

机译:罗氏AMPLICOR HIV-1监测试剂盒光密度序列中的误差对检测结果有效性的影响

获取原文
           

摘要

Specifications for the AMPLICOR HIV-1 MONITOR kit indicate that the results are invalid if the optical densities (ODs) from the PCR-amplified sample that are between 0.1 and 2.3 units are out of sequence. However, among 11,904 assays, results were biased only when ODs of 0.2 to 2.0 units were out of sequence, reducing the rate of invalid results from 3.2 to 0.59%.
机译:AMPLICOR HIV-1 MONITOR试剂盒的规格表明,如果PCR扩增的样品的光密度(OD)在0.1和2.3之间,则该结果无效。但是,在11,904个测定中,仅当OD值从0.2到2.0单位失序时,结果才有偏差,将无效结果的比率从3.2降低到0.59%。