首页> 外文期刊>Journal of Clinical Microbiology >Comparison of Crystal Enteric/Nonfermenter system, API 20E system, and Vitek AutoMicrobic system for identification of gram-negative bacilli.
【24h】

Comparison of Crystal Enteric/Nonfermenter system, API 20E system, and Vitek AutoMicrobic system for identification of gram-negative bacilli.

机译:比较用于鉴定革兰氏阴性杆菌的Crystal Enteric / Nonfermenter系统,API 20E系统和Vitek AutoMicrobic系统。

获取原文
           

摘要

A comparative evaluation of the Crystal Enteric/Nonfermenter system (Crystal; Becton Dickinson, Cockeysville, Md.), API 20E (API; bioMérieux Vitek, Inc., Hazelwood, Mo.), and the Vitek GNI card (Vitek; bioMérieux Vitek) was performed with 512 clinical isolates of gram-negative bacilli, including 381 members of the family Enterobacteriaceae and 131 nonenteric bacilli. With supplemental testing, API, Crystal, and Vitek correctly identified to the genus and species level 505 (98.6%), 489 (95.5%), and 494 (96.5%) of the 512 isolates, respectively. Supplemental testing, as specified by the manufacturer, was required to identify 119 (23.2%), 18 (3.5%), and 5 (1.0%) of the isolates with the three systems, respectively. Of the 381 isolates from the family Enterobacteriaceae, API and Crystal correctly identified 90.3 and 91.6% by 18 to 24 h without supplemental testing, respectively, and Vitek identified 92.4 and 96.1% following 10 and 18 h of incubation, respectively. Of the 131 nonenteric organisms, API and Crystal correctly identified 28.2 and 93.9% by 18 to 24 h without supplemental testing, respectively, and Vitek identified 84.0% by 10 h and 93.9% by 18 h. Errors in identification with each system were infrequent and appeared to be randomly distributed among the genera evaluated. The three systems were comparable in accuracy when either a weighted clinical laboratory profile of organisms or a group of selected isolates in a stress test sample was evaluated (P > 0.05). There were no significant differences between the three systems in their ability to identify either the isolates in the weighted group or those in the stress test (P > 0.05). Crystal compared favorably with API and Vitek, which have established track records in clinical laboratories, and is acceptable for the identification of members of the Enterobacteriaceae and nonenteric bacilli in a clinical microbiology laboratory.
机译:水晶肠溶/非发酵剂系统(水晶;马里兰州科基斯维尔的Becton Dickinson),API 20E(API;密苏里州黑泽尔伍德的BioMérieuxVitek,Inc.)和Vitek GNI卡(Vitek;bioMérieuxVitek)的比较评估用512株革兰氏阴性细菌的临床分离株进行检测,包括381个肠杆菌科成员和131个非肠杆菌。通过补充测试,API,Crystal和Vitek分别正确识别了512个分离株中的505个(98.6%),489个(95.5%)和494个(96.5%)的水平。按照制造商的规定,需要进行补充测试,以分别确定使用这三种系统的119株(23.2%),18株(3.5%)和5株(1.0%)。在没有补充检测的情况下,在18至24小时内,来自肠杆菌科的381株分离物中,API和Crystal分别正确鉴定出90.3和91.6%,在孵育10和18 h后,Vitek分别鉴定出92.4%和96.1%。在131种非肠道生物中,未经补充测试的API和Crystal到18至24 h分别正确地鉴定出28.2%和93.9%,Vitek的10 h和183.9 h分别鉴定出84.0%和93.9%。每个系统的识别错误很少见,并且似乎在所评估的属中随机分布。当评估压力测试样品中的生物的加权临床实验室概况或一组选定的分离株时,这三个系统的准确性可比(P> 0.05)。三种系统在加权组和压力测试中对分离株的鉴定能力没有显着差异(P> 0.05)。 Crystal与在临床实验室中建立了良好记录的API和Vitek相比具有优势,并且可以在临床微生物学实验室中用于鉴定肠杆菌科和非肠杆菌的成员。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号