Evaluation of the MS-2 system for rapid identification of Enterobacteriaceae.

机译:评估MS-2系统以快速鉴定肠杆菌科。

获取原文
   

获取外文期刊封面封底 >>

       

摘要

The precision, accuracy, and other performance characteristics of the MS-2 (Abbott Laboratories, Diagnostic Division, Dallas, Tex.) system for the identification of Enterobacteriaceae were evaluated in a collaborative study involving three clinical laboratories. When identifying 150 unknown, coded organisms, the MS-2 system was 97%, 98%, and 93% accurate, respectively, in three laboratories. The system showed an overall accuracy of 94% when compared with conventional manual tube methods in identifying 1,154 clinical isolates of 26 species of Enterobacteriaceae. Discrepancies between automated and conventional methods were chiefly caused by biochemical variants, especially among Enterobacter species. The MS-2 system was rapid and simple to operate and produced printed results of bacterial identification in 5 h. The cost of disposable components compared favorably with commercial, visually read systems for identifying Enterobacteriaceae.
机译:在一项涉及三个临床实验室的合作研究中,对用于鉴定肠杆菌科的MS-2系统(Abbott实验室,诊断部,得克萨斯州,德克萨斯州)的精度,准确性和其他性能特征进行了评估。在识别150个未知的编码生物时,在三个实验室中,MS-2系统的准确度分别为97%,98%和93%。与传统的手动试管法相比,该系统在鉴定26种肠杆菌科的1154种临床分离株中显示出94%的总体准确性。自动化方法与常规方法之间的差异主要是由生物化学变异引起的,尤其是在肠杆菌属之间。 MS-2系统快速,易于操作,并在5小时内产生了细菌鉴定的打印结果。一次性组件的成本与用于识别肠杆菌科的商业可视读取系统相比具有优势。