Limitations of Vitek GPS-418 Cards in Exact Detection of Vancomycin-Resistant Enterococci with thevanB Genotype

机译:Vitek GPS-418卡在精确检测具有vanB基因型的耐万古霉素肠球菌中的局限性

获取原文
   

获取外文期刊封面封底 >>

       

摘要

The susceptibilities of 20 strains of vancomycin-resistant enterococci (VRE) with the vanB genotype obtained by using Vitek GPS-418 cards were compared with those obtained by the broth dilution method of the National Committee for Clinical Laboratory Standards (NCCLS) (approved standard M7-A4) and with those obtained by the agar screen method using bile esculin azide agar containing 6 μg of vancomycin per ml. Although both the broth dilution and agar screen methods disclosed no discordance, Vitek GPS-418 cards yielded a very major error compared with the results obtained by the reference broth dilution method of the NCCLS. Vitek GPS-418 cards were therefore found to have considerable room for improvement for the accurate detection ofvanB VRE strains.
机译:将使用Vitek GPS-418卡获得的20株具有 vanB 基因型的耐万古霉素肠球菌(VRE)的敏感性与通过美国国家临床实验室标准委员会的肉汤稀释法获得的敏感性进行比较(NCCLS)(批准的标准M7-A4),以及通过使用每毫升含6μg万古霉素的胆汁七叠氮化物琼脂的琼脂筛查方法获得的物质。尽管肉汤稀释法和琼脂筛查方法均未发现不一致之处,但与通过NCCLS的参考肉汤稀释法获得的结果相比,Vitek GPS-418卡产生了非常大的误差。因此,发现Vitek GPS-418卡在准确检测 vanB VRE菌株方面有很大的改进空间。