首页> 外文期刊>Journal of Clinical Microbiology >Detection kinetics for positive blood culture bottles by using the VITAL automated system.
【24h】

Detection kinetics for positive blood culture bottles by using the VITAL automated system.

机译:使用VITAL自动化系统检测阳性血培养瓶的动力学。

获取原文
   

获取外文期刊封面封底 >>

       

摘要

The VITAL system principle is based on homogeneous fluorescence technology. During an 11-month period, a total of 19,706 blood cultures from adult patients hospitalized in various establishments of the Montpellier Teaching Hospital were collected in VITAL bottles, of which 1,939 were declared positive. Only 204 bottles (1.04%) were false positives. The 1,735 true-positive bottles were collected from 130 patients. The final visual control permitted the detection of 10 falsely negative bottles (0.05%), of which 5 contained clinically significant microorganisms from four patients. The kinetics of detection for all microorganisms showed that 66.6% were detected within 24 h, 83.1% within 48 h, 95.5% within 120 h, and 100% within 150 h. No clinical episode would have been missed had a 5-day protocol been used instead of a 7-day protocol. Among the positive bottles, 65.7% were detected by the SLOPE algorithm, 20.1% by the DELTA algorithm, and 14.2% by the THRESHOLD algorithm. This retrospective study of our results shows that a 5-day protocol is sufficient for the detection of septic episodes using the VITAL system.
机译:VITAL系统原理基于均质荧光技术。在11个月内,从VITAL瓶中收集了在蒙彼利埃教学医院各个场所住院的成年患者的19,706份血液培养物,其中1939份被宣布为阳性。只有204瓶(1.04%)为假阳性。从130位患者中收集了1,735个真阳性瓶。最终的视觉控制允许检出10个假阴性瓶(0.05%),其中5个瓶中含有来自4位患者的具有临床意义的微生物。对所有微生物的检测动力学表明,在24小时内检出66.6%,在48小时内检出83.1%,在120小时内检出95.5%,在150小时内检出100%。如果使用5天方案代替7天方案,则不会错过任何临床发作。在阳性瓶中,通过SLOPE算法检测到65.7%,通过DELTA算法检测到20.1%,通过THRESHOLD算法检测到14.2%。这项对我们结果的回顾性研究表明,为期5天的协议足以使用VITAL系统检测败血症发作。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号