Problems affecting performance of the fluorescent treponemal antibody-absorption test for syphilis.

机译:影响梅毒的荧光性端粒抗体吸收测试性能的问题。

获取原文
           

摘要

Immunofluorescent staining of Treponema pallidum was studied to clarify the effect of three factors on the results of the fluorescent treponemal antibody-absorption test: (i) heat inactivation of sera at 56 degrees C for 30 min before testing, (ii) use of multicircle slides, and (iii) tungsten illumination to visualize and assess unstained treponemes on reactive as well as nonreactive smears. It was found that serum inactivation before testing was not necessary for detection of immunoglobin G antibody, but an immunoglobulin M prozone was detected in unheated serum. On multicircle slides, it was demonstrated that a false-positive reaction could be obtained in 30 s at 37 and 25 degrees C if a smear where a nonreactive serum had been placed was crossed by a strongly reactive serum from another circle. Tungsten illumination proved necessary for correct assessment of unstained treponemes on all fluorescent treponemal antibody-aborption test smears, reactive or nonreactive. The possible role of these factors in incorrect fluorescent treponemal antibody-absorption test results is discussed.
机译:研究了梅毒螺旋体的免疫荧光染色,以阐明三个因素对荧光性螺旋体抗体吸收测试结果的影响:(i)在测试前于56摄氏度将血清热灭活30分钟,(ii)使用多环载玻片,以及(iii)钨丝灯照明,以可视化和评估反应性和非反应性涂片上未染色的色氨酸。发现检测前血清灭活对于检测免疫球蛋白G抗体不是必需的,但是在未加热的血清中检测到免疫球蛋白M前带。在多环载玻片上,证明了如果放置了非反应性血清的涂片与来自另一个圆圈的强反应性血清交叉,则在37和25摄氏度的30 s内可获得假阳性反应。事实证明,钨丝灯照明对于正确评估所有反应性或非反应性荧光性视网膜色素上皮抗体吸附试验涂片上未染色的色氨酸是必要的。讨论了这些因素在不正确的荧光性螺旋体抗体吸收测试结果中的可能作用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号