We compared the accuracy of the Gram-stained smear, the acridine orange-stained smear, and the AutoMicrobic system (AMS; Vitek Systems, Inc., Hazelwood, Mo.) in screening for bacteriuria, as detected by conventional cultures. For 1,024 clinical specimens, results with the acridine orange-stained smear and the Gram-stained smear were very similar. When read for the presence of one or more microorganisms or leukocytes per 20 oil immersion fields, both smears were highly sensitive (92.1 and 93.3%, respectively) and moderately specific (70.0 and 61.7%, respectively). Sensitivity was greater for specimens yielding greater than or equal to 10(5) CFU/ml (96.1 and 98.9%, respectively) than for those with 10(3) to 10(4) CFU/ml (81.4 and 78.0%, respectively). Preliminary classification based upon the tinctorial and morphological characteristics of the Gram-stained smear was compatible with culture results in nearly all cases. The accuracy of the Gram-stained smears was not influenced by special cleaning of the microscopic slides, or the level of expertise of the microscopist. For 715 specimens, the sensitivity of the AMS in detecting bacteriuria (91.5%) was very similar to that of the stained smears (92.1 and 95.7%, respectively), but the specificity was significantly higher (83.2% versus 42.6 and 70.0%). Detection of microorganisms by the AMS took an average of 6.3 +/- 3.0 h. These data suggest that the Gram-stained smear is easily interpreted, very sensitive, acceptably specific, and still the optimal rapid method for screening for bacteriuria in most clinical microbiology laboratories.
机译:我们比较了革兰氏染色涂片、,啶橙橙色涂片和AutoMicrobic系统(AMS; Vitek Systems,Inc.,Hazelwood,MO)在筛查细菌尿症中的准确性,这是通过常规培养物检测的。对于1,024个临床标本,with啶橙染色涂片和革兰氏染色涂片的结果非常相似。当每20个油浸场读取一种或多种微生物或白细胞的存在时,两个涂片均高度敏感(分别为92.1和93.3%)和中等特异性(分别为70.0和61.7%)。产量大于或等于10(5)CFU / ml(分别为96.1和98.9%)的标本比灵敏度为10(3)至10(4)CFU / ml(分别为81.4和78.0%)的标本更敏感。在几乎所有情况下,根据革兰氏染色涂片的着色和形态学特征进行的初步分类均与培养结果相符。革兰氏涂片的准确性不受显微镜载玻片的特殊清洁或显微镜专家的专业水平的影响。对于715个标本,AMS检测细菌尿的灵敏度(91.5%)与染色涂片的灵敏度非常相似(分别为92.1和95.7%),但特异性显着更高(83.2%对42.6和70.0%)。通过AMS检测微生物平均需要6.3 +/- 3.0小时。这些数据表明,在大多数临床微生物实验室中,革兰氏染色涂片易于解释,非常敏感,可接受的特异性,仍然是筛选细菌尿的最佳快速方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号