首页> 外文期刊>Journal of Clinical Microbiology >Comparison of enzyme-multiplied immunoassay technique with fluorescence polarization immunoassay for determination of gentamicin and tobramycin levels in serum.
【24h】

Comparison of enzyme-multiplied immunoassay technique with fluorescence polarization immunoassay for determination of gentamicin and tobramycin levels in serum.

机译:酶联免疫法与荧光偏振免疫法测定血清中庆大霉素和妥布霉素含量的比较。

获取原文
   

获取外文期刊封面封底 >>

       

摘要

We assayed serum gentamicin and tobramycin specimens by the enzyme multiplied immunoassay technique (Syva EMIT) and the fluorescence polarization immunoassay (Abbott TDx). When interassay and intraassay control samples were evaluated, both methods gave an overall coefficient of variation of less than +/- 10%. Using patient serum samples, we obtained excellent correlation with both methods in the assay of gentamicin (correlation coefficient, 0.985) and tobramycin (correlation coefficient, 0.982).
机译:我们通过酶多重免疫测定技术(Syva EMIT)和荧光偏振免疫测定(Abbott TDx)测定了血清庆大霉素和妥布霉素样品。当评估批间和批内对照样品时,两种方法的总变异系数均小于+/- 10%。使用患者血清样本,我们在庆大霉素(相关系数为0.985)和妥布霉素(相关系数为0.982)的测定中获得了与两种方法的极佳相关性。