An immunofluorescence-native deoxyribonucleic acid (nDNA) antibody test system (Zeus Scientific, Inc.) was compared with a radioimmunoassay procedure (FARR assay) for detecting anti-nDNA antibodies in human serum. Double-blind studies of split samples obtained from 236 patients showed an 80% correlation between the immunofluorescence-nDNA antibody test system and the radioimmunoassay procedure. Studies of sera from patients with known diagnoses showed positive nDNA antibody findings in biopsy-prove n systemic lupus erythematosus only. The immunofluorescence-nDNA antibody test system provides a reliable, simple, and economically feasible alternative method for detecting nDNA antibodies that can be employed in any laboratory equipped with a fluorescence microscope.
机译:将免疫荧光天然脱氧核糖核酸(nDNA)抗体测试系统(Zeus Scientific,Inc.)与放射免疫测定法(FARR测定)进行了比较,以检测人血清中的抗nDNA抗体。对来自236位患者的分割样品进行的双盲研究表明,免疫荧光-nDNA抗体测试系统与放射免疫测定方法之间的相关度为80%。对已知诊断患者的血清进行的研究表明,仅在活检证实的系统性红斑狼疮中,nDNA抗体阳性。免疫荧光-nDNA抗体测试系统为检测nDNA抗体提供了可靠,简单且经济可行的替代方法,可在配备荧光显微镜的任何实验室中使用。