首页> 外文期刊>Journal of Light & Visual Environment >A measuring technique for spectral distribution of light sources using an acoustooptic tunable filter
【24h】

A measuring technique for spectral distribution of light sources using an acoustooptic tunable filter

机译:一种使用声光可调滤光片的光谱分布测量技术

获取原文
           

摘要

References(20) This paper describes the characteristics of spectrometer and the measuring technique for spectral distribution in the visible wavelength region using an Acoustooptic Tunable Filter (AOTF) employing a TeO2 crystal. The contents of this paper are summarized as follows: (1) The polarization properties of the diffracted lights of an AOTF are linear polarization. (2) The diffracted light distribution can be approximated to the Gaussian distribution. (3) The resolving power of the wavelength of the diffracted light is smaller than 3 nm. (4) The wavelength accuracy of the diffracted light is about 0.002 nm/kHz.(5) The error of the spectral distribution is less than ±1% over the range of 380∼500 nm and ±0.5% at the longer than 500 nm. (6) The wavelength and spectral transmittance of an AOTF can be controlled electronically without any mechanical means.
机译:参考文献(20)本文介绍了使用TeO2晶体的声光可调滤光片(AOTF)在可见光波长区域内的光谱仪特性和光谱分布的测量技术。本文的内容概括如下:(1)AOTF衍射光的偏振特性是线性偏振。 (2)衍射光分布可以近似为高斯分布。 (3)衍射光的波长的分辨力小于3nm。 (4)衍射光的波长精度约为0.002 nm / kHz。(5)光谱分布的误差在380〜500 nm范围内小于±1%,在大于500 nm范围内小于±0.5% 。 (6)AOTF的波长和光谱透射率可以通过电子方式进行控制,而无需任何机械手段。