Reduction of applicator displacement in MR/CT-guided cervical cancer HDR brachytherapy by the use of patient hover transport system

机译:通过使用患者悬停运输系统减少MR / CT引导的子宫颈癌HDR近距离治疗中的涂药器移位

获取原文
           

摘要

Purpose To quantify the reduction of relative displacement between the implanted intracavitary applicator and the patient bony anatomy, due to the use of a hover transport system during the patient transports between the imaging table and the treatment table. Material and methods The displacement of the applicator inside the patient was measured by comparing the distance between the tip of the tandem and the pubic bone on X-ray radiography images taken before and after moving a patient to magnetic resonance/computed tomography imaging. Displacements were evaluated for 27 fractions of treatment using hover transport and 185 fractions of treatment using manual transport. Results The use of hover transport system reduced the percentage of fractions with displacements greater than 5 mm from 22.7% to 7.4%. The reduction of applicator displacement using hover transport is statistically significant, compared to the manual transport method (p-value 0.0086; mean displacement 3.41 mm [95% CI: 2.96-3.97] for manual transport, and 2.27 mm [95% CI: 1.71-2.97] for hover transport fractions). Conclusions This study indicates that the hover transport system is effectively reducing displacement between tandem and patient bony anatomy during patient transports. The potential improvement in dosimetric accuracy due to this reduction warrants further study.
机译:目的为了量化由于在成像台和治疗台之间进行患者运输期间使用悬停运输系统而导致的腔内施药器与患者骨解剖结构之间相对位移的减少。材料和方法通过比较在将患者转移到磁共振/计算机断层扫描成像之前和之后所拍摄的X射线射线照相图像上,比较串联器尖端和耻骨之间的距离来测量施药器在患者体内的位移。使用悬停运输评估位移的27个处理部分,使用手动运输评估位移的185个部分。结果悬停运输系统的使用将位移大于5 mm的馏分百分比从22.7%降低到7.4%。与手动运输方法相比,使用悬停运输减少施药器位移具有统计学意义(p值0.0086;手动运输平均位移3.41毫米[95%CI:2.96-3.97],2.27毫米[95%CI:1.71] -2.97](用于悬停运输分数)。结论这项研究表明,悬停运输系统有效地减少了患者运输过程中串联和患者骨骼解剖之间的移位。由于这种减少,剂量学准确性的潜在提高值得进一步研究。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号