Comparison of the COBAS/Ampliprep Taqman and Amplicor HIV-1 monitor tests in Lagos, Nigeria

机译:尼日利亚拉各斯的COBAS / Ampliprep Taqman和Amplicor HIV-1监测仪测试的比较

获取原文
   

获取外文期刊封面封底 >>

       

摘要

Background: The use of real-time Polymerase chain reaction (PCR) technology options is increasing in resource-limited settings because they are faster, improve assay sensitivity,have higher throughput, larger dynamic ranges and reduced rates of contamination. In 2010, UNAIDS ranked Nigeria as the second highest population of people living with HIV and AIDS (2.98 million people) in the world.Objective: The objective of this study was to compare the analytical performances of the Amplicor HIV-1 Monitor (version 1.5) and the COBAS Ampliprep/Taqman (version 2.0) usedin monitoring HIV disease progression in HIV-infected individuals.Method: In a cross-sectional study, HIV-1 RNA values obtained with the Amplicor HIV-1 monitor version 1.5 were compared with those of the COBAS/Ampliprep TaqMan HIV-1version 2.0 in a routine clinical setting. Between May and November 2011, 176 plasma samples collected were analysed in parallel using both techniques. Data analysis was done using statgraphics Centurion XVI and Medcalc version 12.0.Result: The correlation coefficient for the two assays was 0.83 and the level of agreement using a Bland–Altman plot was 94.2%.Conclusion: These findings suggest that the results from the two methods were comparable, hence the COBAS/Ampliprep Taqman version 2.0 is recommended for high-volume laboratories.
机译:背景:在资源有限的环境中,实时聚合酶链反应(PCR)技术选项的使用正在增加,因为它们更快,提高测定灵敏度,具有更高的通量,更大的动态范围并降低了污染率。 2010年,联合国艾滋病规划署将尼日利亚列为世界第二大艾滋病毒/艾滋病感染者(298万)。目的:本研究的目的是比较Amplicor HIV-1监测仪(1.5版)的分析性能)和COBAS Ampliprep / Taqman(2.0版)用于监测HIV感染者的HIV疾病进展。方法:在一项横断面研究中,将使用Amplicor HIV-1监测器1.5版获得的HIV-1 RNA值与那些进行比较。常规临床环境中对COBAS / Ampliprep TaqMan HIV-1version 2.0的检测。在2011年5月至2011年11月之间,使用这两种技术并行分析了收集的176个血浆样品。使用统计数据Centurion XVI和Medcalc版本12.0进行数据分析。结果:两种测定的相关系数为0.83,使用Bland-Altman图的一致性为94.2%。结论:这些发现表明,两种测定的结果这些方法具有可比性,因此建议将COBAS / Ampliprep Taqman 2.0版用于大容量实验室。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号