Malposition of Cage in Minimally Invasive Oblique Lumbar Interbody Fusion

机译:微创斜向腰椎椎间融合器笼的位置不正确

获取原文
           

摘要

Introduction. Minimally invasive oblique lumbar interbody fusion is one of the novel lateral lumbar interbody fusion techniques for which the successful early results have been reported. However, new complications were increasingly reported from ongoing studies. Case Presentation. We report a case of an unusual complication of minimally invasive oblique lumbar interbody fusion associated with contralateral nerve root compression due to deep and posterior position of polyetheretherketone cage and discussion of the operating technique for repositioning polyetheretherketone cage. Conclusion. Malposition of polyetheretherketone cage can cause contralateral nerve root compression and neurological complication. The surgical technique to proper pull the polyetheretherketone cage back into the acceptable position should be considered and well prepared.
机译:介绍。微创倾斜腰椎椎间融合术是新型的外侧腰椎椎间融合术之一,其成功的早期结果已有报道。但是,正在进行的研究越来越多地报道了新的并发症。案例介绍。我们报告一例因聚醚醚酮笼的深部和后部位置而引起对侧神经根受压而引起的微创斜腰椎间融合术的异常并发症,并讨论了重新定位聚醚醚酮笼的操作技术。结论。聚醚醚酮笼的位置不当会引起对侧神经根受压和神经系统并发症。应考虑并做好充分准备将聚醚醚酮笼适当拉回可接受位置的手术技术。