BackgroundWe studied the worst-case radiated radiofrequency (RF) susceptibility of automated external defibrillators (AEDs) based on the electromagnetic compatibility (EMC) requirements of a current standard for cardiac defibrillators, IEC 60601-2-4. Square wave modulation was used to mimic cardiac physiological frequencies of 1 - 3 Hz. Deviations from the IEC standard were a lower frequency limit of 30 MHz to explore frequencies where the patient-connected leads could resonate. Also testing up to 20 V/m was performed. We tested AEDs with ventricular fibrillation (V-Fib) and normal sinus rhythm signals on the patient leads to enable testing for false negatives (inappropriate "no shock advised" by the AED).
机译:背景我们基于现行心脏除颤器标准IEC 60601-2-4的电磁兼容性(EMC)要求,研究了自动体外除颤器(AED)在最坏情况下的辐射射频(RF)敏感性。方波调制用于模拟1-3 Hz的心脏生理频率。与IEC标准的差异是将频率下限限制为30 MHz,以探索患者连接的导线可能产生共振的频率。还进行了高达20 V / m的测试。我们用心室颤动(V-Fib)和正常窦性心律信号对AED进行了测试,从而可以测试假阴性(AED提出了“不建议休克”)。