>Iron isotopic analyses of geological reference materials on MC-ICP-MS with instrumental mass bias corrected by three independent methods
Iron isotopic analyses of geological reference materials on MC-ICP-MS with instrumental mass bias corrected by three independent methods
Here we report iron(Fe) isotopic data of three pure Fe solution standards(IRMM-014, GSB Fe, and NIST3126a) and five widely used geological reference materials(RMs) from the United States Geological Survey and Geological Survey of Japan obtained on a Neptune Plus multi-collector–inductively coupled plasma–mass spectrometer(MC-ICP-MS) in our laboratory over the past 3 years. The instrumental mass bias was corrected by three independent methods: sample-standard bracketing(SSB),Ni doping + SSB, and ^(57)Fe–^(58)Fe double spike + SSB.Measurements reveal that both the Ni doping and double spike methods helped calibrate short-term fluctuations in mass bias. Collectively, almost all measurements of RMs yielded δ^(56)Fe within ± 0.05 of recommended values,provided that each sample was measured four times on MC-ICP-MS. For the first time, new recommended values for NIST SRM3126a are reported(δ^(56)Fe = 0.363 ± 0.006,2SE, 95% CI; and δ^(57)Fe = 0.534 ± 0.010, 2SE).